Skip to main content
Discontinued

Interface 5000P

Replaced by Interface 5030P

Key features include:

  • Maximum Applied Potential - ±6 V
  • Maximum Current - +/- 5A
  • Measure both half cells simultaneously
  • EIS - 10 μHz - 20 kHz

This page is kept as archival reference for existing owners — it is not available for purchase.

Interface 5000P

System Highlights

Interface 5000P

The Gamry Interface 5000P is designed for people doing cell-level testing of various energy storage and conversion devices.  It even includes capabilities for monitoring both half cell voltages in addition to the whole cell voltage during an experiment. Imagine the time savings in being able to characterize both half cells and the full cell in the same experiment.  

Characterizing Both Half Cells

The Interface 5000P has the ability to monitor both half cells during experiments. Two electrometers measure the potential of the anode and cathode in addition to the full cell voltage when used with an embedded reference electrode.

dual electrometers for monitoring both anode and cathode, in addition to the full
(Click image to Enlarge)

Upgradeabilty

The Interface 5000P also comes in a more advanced model, the Interface 5000E.  This model provides full capabilities for all application areas.  An in the field upgrade is available to upgrade an Interface 5000P to an Interface 5000E.  Contact us for details regarding upgrades.

Need More Capabilities?  See the Interface 5000E

More Details

Below are additional details regarding the capabilities of the Interface 5000P potentiostat/galvanostat/ZRA. Each bullet point contains a list of the type of techniques available for the instrument to run.

Other details

  • 2, 3, 4 and 5 electrode measurements
  • Measure both half cell voltages in addition to the full cell voltage
  • Electrical Isolation
    • Floating instrument: use with autoclaves, mechanical stress apparatus, or pipeline probes.
  • Portable
    • Size of a chemistry textbook, weighing only 2.5 kgs. Easy USB 2.0 connection to a Windows computer.
  • Built-In EIS
    • On-board DDS to perform EIS from 10 µHz to 20 kHz. 
  • DSP (Digital Signal Processing) Mode
    • Oversamples for improved signal-to-noise and accurate capacitance measurements.
  • Current Interrupt iR Compensation
    • Gamry potentiostats and their controlling software use control loop algorithms to accurately measure and correct for uncompensated resistance.
  • Auxiliary I/O
    • Control additional equipment via additional I/O interfaces: external signal input, analog voltage output, analog current output, auxiliary A/D input, and digital I/O connector.
    • RTD temperature probe measurement
  • Warranty
    • Protected by 2-year factory service warranty.

Specifications

System
Cell Connections2, 3, 4 or 5
Maximum Current±5 A
Current Ranges6 (50 μA - 5 A)
Current Ranges (including internal gain)8
Minimum Current Resolution150 pA
Maximum Applied Potential±6 V
Rise Time<1 μs
Minimum Time Base10 μs
Noise and Ripple<20 μV rms
Control Amplifier
Compliance Voltage±6.5V
Output Current>±5 A
Speed Settings5
Unity Gain Bandwidth1050, 250, 43, 4.4, 0.5 kHz
EIS Measurement
EIS10 μHz - 20 kHz
Current AC Amplitude5 A maximum
Electrometer
Input Impedance>1012 Ω || <2 pF
Input Current (typical)<25 pA
Bandwidth>12 MHz at -3 dB
CMRR>98 dB (10 kHz)>88dB (100 kHz)
Potential
Applied Accuracy±1 mV ±0.2% of setting
Applied Resolution200 μV, 50 μV, 12.5 μV/bit
Measured Accuracy±0.5 mV ±0.2% of reading
Measured Resolution200 μV, 20 μV, 2 μV/bit
Current
Applied/Measured Accuracy±25 pA ±0.05% of range ±0.2% of value
Applied/Measured Resolution0.0033% full-scale/bit
Bandwidth>5 MHz (5 mA)