Interface 1000™

The Interface 1000 is a research grade Potentiostat/Galvanostat/ZRA for use in general electrochemistry applications. It is ideal for corrosion measurements, battery testing, sensor development, and physical electrochemistry. It is the perfect blend of performance and value.

Overview

Flexible
With 9 current ranges and 3 gain stages, you have 11 decades of current measurement at your command. From corrosion to batteries, sensors to supercaps, the Interface 1000 has got you covered. 

Affordable
By combining state of the art electronics with expert design, the Interface 1000 is quite simply the best value in potentiostats. 
Powerful
Cycle your batteries. Polarize those large electrodes. The Interface 1000 sports a full 1 Ampere of current up to the compliance limit of +/-20 V. 
Floating
Like all Gamry Potentiostats, the Interface 1000 was designed from the ground up for true floating capability. Whether you need to measure earth grounded working electrodes, or multiple electrodes in a shared electrolyte, the Interface 1000 is up to the task. 
Low Noise
The Interface 1000 continues the Gamry tradition with an exceptional < 20 uV noise specification. Low noise = better measurements. 
Impedance on Board
No external modules. No expensive add-ons. The Interface 1000 acquires accurate impedance measurements up to 1 MHz. 
Configurable Multichannel
Why sacrifice convenience when you need more than one potentiostat? The Interface 1000 lets you control where each channel is located. Put channel 1 in your hood, channel 2 in your glovebox, and dock channels 3-8 on the bench. You wanted flexibility, we delivered!

Affordable
By combining state of the art electronics with expert design, the Interface 1000 is quite simply the best value in potentiostats.

Powerful
Cycle your batteries. Polarize those large electrodes. The Interface 1000 sports a full 1 Ampere of current up to the compliance limit of +/-20 V.

Floating
Like all Gamry Potentiostats, the Interface 1000 was designed from the ground up for true floating capability. Whether you need to measure earth grounded working electrodes, or multiple electrodes in a shared electrolyte, the Interface 1000 is up to the task.

Low Noise
The Interface 1000 continues the Gamry tradition with an exceptional < 20 μV noise specification. Low noise = better measurements.

Impedance on Board
No external modules. No expensive add-ons. The Interface 1000 acquires accurate impedance measurements up to 1 MHz.

Configurable Multichannel
Why sacrifice convenience when you need more than one potentiostat? The Interface 1000 lets you control where each channel is located. Put channel 1 in your hood, channel 2 in your glovebox, and dock channels 3-8 on the bench. You wanted flexibility, we delivered!

Accuracy Contour Plot

The Accuracy Contour Plot provides a detailed look at the performance you can expect from your instrument in real-world situations. The results include the cell cable.

Interface 1000 Accuracy Contour Plot

 
Interface 1000
Potentiostat YES
Galvanostat YES
Zero Resistance Ammeter YES
Cell Connections 2, 3, or 4
Floating (Isolated from Earth)  YES
   

WEIGHT

2.4 kg

DIMENSIONS

24 (W) x 6 (H) x 27 (D) cm
   

SYSTEM

 
Max Current ±1 A
Current Ranges 9 (10 nA - 1 A)
Current Ranges (w/Internal
Gain applied)
11 (100 pA - 1 A)
Min Voltage Resolution 1 μV
Min Current Resolution 3.3 fA
Max Applied Potential ±12 V
Rise Time TBD
Noise and Ripple <20 μVrms
Min Time Base 10 μs
Max Time Base 715 s
Min Potential Step 12.5 μV
Analog/Digital Converter 16 bit
   

EIS MEASUREMENT

 
Frequency Range 10 μHz - 1 MHz
EIS Accuracy See Accuracy
Contour Plot
Max AC Amplitude 2.33 Vrms
Min AC Amplitude 17.8 μVrms
   

CONTROL AMP

 
Compliance Voltage ±20 V
Output Current >±1 A
Speed Settings 5
Unity Gain Bandwidth  980, 260, 40, 4, 0.4 kHz
   

ELECTROMETER

 
Input Impedance >1012 Ω
Input Current <20 pA
Bandwidth (-3dB) (typical) >10 MHz
Common Mode Rejection Ratio >80 dB (10 kHz)
>60 dB (1 MHz)
   

APPLIED POTENTIAL

 
Accuracy ±1 mV ±0.2% of setting
Resolution 12.5 μV, 50 μV, 200 μV/bit
Drift <20 μV/°C
Potential Scan Range ±0.4 V, ±1.6 V, ±6.4 V
   

MEASURED POTENTIAL

 
Accuracy ±1 mV ±0.3% of reading
Full-Scale Ranges ±12 V, ±3 V, ±300 mV, ±30 mV
Resolution 400 μV, 100 μV, 10 μV, 1 μV/bit
Offset Range ±12 V, ±3 V
   

APPLIED CURRENT

 
Accuracy ±5 pA ±0.3% of range
Resolution 0.0033% full-scale/bit
   

MEASURED CURRENT

 
Accuracy ±5 pA ±0.3% of range
Resolution 0.0033% full-scale/bit
Bandwidth (-3 dB)
NOTE: Bandwidth is current 
         range dependent.
> 10 MHz (1A - 100 μA)
> 1.5 MHz (10 μA)
> 0.15 MHz (1 μA)
Stability Settings 3
Post Offset Gain 1X, 10X, 100X
Offset Range ±1X full-scale
   

iR COMPENSATION

 
Mode Current Interrupt
Min Interrupt Time 33 μs
Max Interrupt Time 715 s