info@gamry.com | 1-877-367-4267
The Interface 1000 is a research grade Potentiostat/Galvanostat/ZRA for use in general electrochemistry applications. It is ideal for corrosion measurements, battery testing, sensor development, and physical electrochemistry. It is the perfect blend of performance and value.
Flexible
With 9 current ranges and 3 gain stages, you have 11 decades of current measurement at your command. From corrosion to batteries, sensors to supercaps, the Interface 1000 has got you covered.
Affordable
By combining state of the art electronics with expert design, the Interface 1000 is quite simply the best value in potentiostats.
Powerful
Cycle your batteries. Polarize those large electrodes. The Interface 1000 sports a full 1 Ampere of current up to the compliance limit of +/-20 V.
Floating
Like all Gamry Potentiostats, the Interface 1000 was designed from the ground up for true floating capability. Whether you need to measure earth grounded working electrodes, or multiple electrodes in a shared electrolyte, the Interface 1000 is up to the task.
Low Noise
The Interface 1000 continues the Gamry tradition with an exceptional < 20 μV noise specification. Low noise = better measurements.
Impedance on Board
No external modules. No expensive add-ons. The Interface 1000 acquires accurate impedance measurements up to 1 MHz.
Configurable Multichannel
Why sacrifice convenience when you need more than one potentiostat? The Interface 1000 lets you control where each channel is located. Put channel 1 in your hood, channel 2 in your glovebox, and dock channels 3-8 on the bench. You wanted flexibility, we delivered!
The Accuracy Contour Plot provides a detailed look at the performance you can expect from your instrument in real-world situations. The results include the cell cable.

Interface 1000 |
|
| Potentiostat | YES |
| Galvanostat | YES |
| Zero Resistance Ammeter | YES |
| Cell Connections | 2, 3, or 4 |
| Floating (Isolated from Earth) | YES |
WEIGHT |
2.4 kg |
DIMENSIONS |
24 (W) x 6 (H) x 27 (D) cm |
SYSTEM |
|
| Max Current | ±1 A |
| Current Ranges | 9 (10 nA - 1 A) |
| Current Ranges (w/Internal Gain applied) |
11 (100 pA - 1 A) |
| Min Voltage Resolution | 1 μV |
| Min Current Resolution | 3.3 fA |
| Max Applied Potential | ±12 V |
| Rise Time | TBD |
| Noise and Ripple | <20 μVrms |
| Min Time Base | 10 μs |
| Max Time Base | 715 s |
| Min Potential Step | 12.5 μV |
| Analog/Digital Converter | 16 bit |
EIS MEASUREMENT |
|
| Frequency Range | 10 μHz - 1 MHz |
| EIS Accuracy | See Accuracy Contour Plot |
| Max AC Amplitude | 2.33 Vrms |
| Min AC Amplitude | 17.8 μVrms |
CONTROL AMP |
|
| Compliance Voltage | ±20 V |
| Output Current | >±1 A |
| Speed Settings | 5 |
| Unity Gain Bandwidth | 980, 260, 40, 4, 0.4 kHz |
ELECTROMETER |
|
| Input Impedance | >1012 Ω |
| Input Current | <20 pA |
| Bandwidth (-3dB) (typical) | >10 MHz |
| Common Mode Rejection Ratio | >80 dB (10 kHz) >60 dB (1 MHz) |
APPLIED POTENTIAL |
|
| Accuracy | ±1 mV ±0.2% of setting |
| Resolution | 12.5 μV, 50 μV, 200 μV/bit |
| Drift | <20 μV/°C |
| Potential Scan Range | ±0.4 V, ±1.6 V, ±6.4 V |
MEASURED POTENTIAL |
|
| Accuracy | ±1 mV ±0.3% of reading |
| Full-Scale Ranges | ±12 V, ±3 V, ±300 mV, ±30 mV |
| Resolution | 400 μV, 100 μV, 10 μV, 1 μV/bit |
| Offset Range | ±12 V, ±3 V |
APPLIED CURRENT |
|
| Accuracy | ±5 pA ±0.3% of range |
| Resolution | 0.0033% full-scale/bit |
MEASURED CURRENT |
|
| Accuracy | ±5 pA ±0.3% of range |
| Resolution | 0.0033% full-scale/bit |
| Bandwidth (-3 dB) NOTE: Bandwidth is current range dependent. |
> 10 MHz (1A - 100 μA) > 1.5 MHz (10 μA) > 0.15 MHz (1 μA) |
| Stability Settings | 3 |
| Post Offset Gain | 1X, 10X, 100X |
| Offset Range | ±1X full-scale |
iR COMPENSATION |
|
| Mode | Current Interrupt |
| Min Interrupt Time | 33 μs |
| Max Interrupt Time | 715 s |