Run the Cyclic Voltammetry Electrochemical Energy standard technique by selecting Experiment>Electrochemical Energy>Cyclic Voltammetry from the Framework™ menu bar.
This initiates the following sequence of events:
The script conditions the electrode if Conditioning was specified in the Setup. Conditioning is done by applying a fixed potential for a defined time. A plot of current versus time is displayed during Conditioning.
Finally an actual scan occurs. The potential of the cell is set to Initial E. The potential is then swept from the Initial E to Scan Limit 1, then to Scan limit 2 and then to Final E. If Scan Llimit 2 equals Final E then the scan stops at that value. Current readings at fixed voltage-intervals are taken during the sweep. The voltage interval between steps is defined as the Resolution in the Setup dialog box.
If the number of Cycles > 1, the potential repeats the sweep for n cycles.
The sweep is actually a staircase ramp. The sample is potentiostatted at the Initial E, a delay of one sample period occurs (sample period = 1/Scan Rate × Resolution), and a reading of the current is taken. The potential is then stepped by a few mV as defined by Resolution, a delay of one Sample Period occurs, and the next current reading is taken. Stepping the potential, delaying and acquiring data points continues until the potential equals the Final E.
If you selected Autoranging in Setup, at each point the current range is automatically switched to the optimal range for the measured cell current. If you selected Positive Feedback IR Compensation, all data are continuously corrected for IR-drop. If you selected Current Interrupt IR Compensation, each potential is corrected for the measured IR-drop of the preceding point.
A plot of I vs. E is displayed during the scan.