Search this site
English
Contact Us

Low-impedance EIS at its Limits: Gamry Reference 30k Booster

Introduction

This Application Note describes measurement of very small impedances using Electrochemical Impedance Spectroscopy (EIS). We recorded EIS spectra of precision current shunts to demonstrate the ability of a Gamry Instruments system to accurately measure impedances less than 100 µΩ. This note discusses the performance limits in systems and describes background subtraction to extend performance. 

We assume the reader has a basic understanding of EIS and its applications.  If you need a refresher, please read Gamry’s Applications Note: “Basics of Electrochemical Impedance Spectroscopy”.

Background

Accurate measurement of low impedance is of greatest importance in electrochemical energy devices (batteries, fuel cells, super-capacitors, photovoltaic cells). Series resistance within these devices causes unwanted losses in power and energy. 

Sources for series resistance include:

  • Contact resistance,
  • Resistance of metallic current-collectors,
  • Poorly conductive active materials, and
  • Resistance in electrolytes and separators. 

Standard practice in electrochemical energy devices sums all series resistive terms into a single term called “Equivalent Series Resistance” or ESR. Batteries, fuel cells, and super-capacitors all require very low ESR to maximize power, energy efficiency, and device lifetime.   

EIS tests on electrochemical energy devices reveal more than just ESR. EIS can also measure capacitance, electrode porosity, and kinetics of electrochemical reactions. A device’s EIS spectrum typically changes during slow capacity fade and when the device fails.

These changes can be used as a diagnostic tool to determine the failure mechanism. Accuracy in low-impedance EIS is a must for all these measurements.

Low-impedance measurements, i.e., below 1 Ω, are most accurately done galvanostatically: an AC current is applied to an electrochemical cell and the AC voltage across the cell is measured. Accurate, very low impedance measurements—below 1 mΩ—are particularly challenging. 

For example, assume we want a measurement accurate to within ± 1% of a 100 µΩ device using AC stimulation at 1 Arms. The voltage across the device is 100 µVrms so our measurement error must be less than 1 µVrms. Noise and drift on voltage signals can easily add up to tens of µV, so care and special techniques are needed to resolve one µV. 

Higher current makes the voltage bigger so the measurement is easier. Unfortunately higher current often has these limitations:

  • Some systems are damaged by high current, 
  • High-current, high-power instrumentation can be quite expensive,
  • Bandwidth often falls as current increases,
  • Finding and connecting cables can get difficult.  

System and Cable Limitations

Impedances between 1 Ω and 1 kΩ generally can be measured easily and accurately. Details of the test setup and test method have little influence on the results. The measured impedance is usually free from artifacts generated by the test system. 

As the measured impedance drops, the contribution of the measurement system (potentiostat plus cables and connections) becomes a larger factor. There are three ways that most systems have limits to low-impedance measurements: 

1)     Instrument noise limits the smallest voltage that can be accurately measured. EIS uses a phase-sensitive detection scheme to minimize the bandwidth of noise that can affect the measurement, but this can only lower noise, not eliminate it. Noise in the measured voltage generally causes noise in the reported impedance.  

2)     Real-world instrumentation always has some leakage of the applied signal into the measured signal. Explanation of the causes of the leakage is beyond the scope of this note. When measuring very small impedances, the voltage signal that represents cell current can be 106 times larger than the sensed voltage signal.

Signal-leakage error often models as a resistance in series with the “true” cell impedance. The apparent resistance value can be negative!

3)    Additionally, the cell current flowing through the cables connecting the instrument to the cell generates a magnetic field that can be picked up in the voltage sensing leads. This effect, called mutual inductance, is a function of frequency. Mutual-inductance error is generally insignificant at low frequency, and totally dominates the measurement at high frequency. The impedance error from mutual inductance generally models as an apparent inductance in series with the “true” cell impedance.  

Moving the cell leads changes mutual-inductance errors. The apparent inductance value can be negative.

Experimental Details

All spectra in this note were recorded using Gamry’s Electrochemical Impedance Spectroscopy Software. The Gamry hardware was a Reference 3000 Potentiostat/Galvanostat/ZRA connected to a Reference 30k Booster. Standard cell cables were used in a 4‑terminal arrangement. The current shunts were Empro Models HA-400-50, FL‑2000-50 and FL-3000-50. The latter two shunts required special connections described at the end of this note.

The Li-Ion battery we tested was a Gaia HP-602050 rated for 45 Ah. The super-capacitor we tested was a Maxwell BCAP0650 P270 rated for 650 F. Both tests used an intermediate state-of-charge on the device.

All shunt tests were done with an applied current of 20 Arms. The battery and super-capacitor spectra used an excitation of 3 Arms. Shorted lead tests were measured at both 20 Arms and 3 Arms so as to best compare with the shunt and device spectra.

Shorted Lead

An EIS system’s minimum impedance characteristic can be measured via a shorted lead test. The cell leads are shorted together and a galvanostatic EIS spectrum is recorded. The excitation current is usually the maximum current allowed in the system. 

This test can be very tricky for an EIS system capable of measuring very low impedance. The tester must carefully place the cell leads to avoid errors. One often misunderstood problem in shorted lead tests is the finite resistance of even the most conductive metals. The resistance of a cubic block of copper 1 cm on each side is 1.6 µΩ. While this is a very small resistance, it is definitely not zero and can be measured using a good EIS system.

Figure 1 shows the arrangement of cell leads Gamry uses to record accurate shorted-lead spectra. The current-carrying leads (red and green) are both bolted to a copper block. The voltage sense leads (blue and white) are connected to a post with no current flow through it.   The resistance of the copper block is not measured when you use this lead arrangement. 


shorted lead test fixture

Figure 1.  Shorted-lead test fixture. Sense leads are connected at top left, current leads at bottom right.

The sense leads should be twisted together to minimize loop area for magnetic coupling. The sense leads should be kept far away from the current-carrying leads. This minimizes mutual-inductance errors. The mutual inductance effect is not very reproducible from one set-up to the next. 

The arrangement of the cell leads while testing a real device is likely to be different than the arrangement during the shorted-lead test. The mutual inductance error will therefore be different.

Shorted Lead Results

Figure 2 is a Bode plot of the shorted-lead spectrum we measured using a boosted system and an excitation current of 20 Arms. The solid line is a least-squares fit of these data to a series-RL model. 

shorted lead spectrum

Figure 2 – Shorted-lead spectrum for Reference 3000 with Reference 30k Booster.

The fit values are –144.0 ± 0.5 nΩ and –10.82 ± 0.07 pH. As mentioned above, signal leakage and mutual inductance can cause negative apparent resistance and inductance. 

Also note that the phase data look quite bad. This is simply an artifact of the negative resistance, making current and voltage 180° out of phase, and noise randomly making that 180° phase either positive or negative. 

Surrogate Measurements

In some cases, a cell surrogate gives a better baseline spectrum than the shorted-lead experiment described above. A cell surrogate is a very-low-impedance model of an electrochemical device with the same geometry as the real device. A good example is a fuel cell in which a Membrane Electrode Assembly (MEA) is sandwiched between metal plates. The fuel-cell surrogate could use a copper foil in place of the MEA. 

In general, a cell surrogate models inductance errors very well because the cell leads can be kept in the same geometry for both the surrogate and the test cell. It is not as good in modeling resistance errors unless the surrogate’s resistance is much smaller than the cell’s resistance.

Cell-surrogate measurements were not made on the shunts.

Background Subtraction

Background subtraction is very useful in correcting for low impedance measurement errors. Of course, it is only valid for stable systems.  

Gamry’s software allows for subtraction of circuit elements and complete spectra. Low-impedance errors generally model as impedance in series with a cell under test, so always use background-subtraction of a series element or spectrum for background correction.  

We point out two cautions:

1)       The shorted-lead spectrum and the surrogate spectrum are each good at modeling one aspect of system behavior but much poorer at modeling other aspects. Spectrum subtraction can have problems. The best correction may be series subtraction of a resistance obtained by modeling a shorted-lead spectrum, and series subtraction of an inductor obtained from modeling a surrogate spectrum. 

2)       Be very careful if the impedance of the system is similar to the background. Subtraction of two large measured numbers rarely leads to an accurate smaller number. 

Measuring Current Shunts

Current shunts are precision low-resistance devices designed for measurement of very high currents. Current shunts generally are designed to produce a 50 mV drop at their rated current level. The shunts used in this note are rated to have error less than ± 0.25%. You can buy them for less than US $100. 

Impedance spectra were recorded on shunts with resistance of 16.67 µΩ, 25 µΩ, and 125 µΩ. They were specified for 50 mV drop at 3000 A, 2000 A and 400 A respectively. The shunts are shown in Figure 3.

measured shunts

Figure 3 – . Measured shunts and the shorted-lead test fixture.

125 μΩ Shunt

The first shunt tested was rated for 50 mV at 400 A, meaning a resistance of 125.0 ± 0.3 µΩ. This small shunt has a simple setup with only one point of contact on each side for the current connections. Figure 4 shows a Bode plot of this shunt’s impedance spectrum.

Low-Impedance-EIS-at-its-Limits-Ref30k-Booster.pdf 3.jpg

Figure 4 – Impedance spectrum of 125 µΩ shunt.

The flat portion of the curve corresponds to 125.0 ± 0.3 µΩ at 0° phase. All measurements up to 100 Hz are flat to within 1% magnitude and within 2° phase.

25 μΩ Shunt

The second shunt tested was rated for 50 mV at 2000 A or 25.00 ± 0.06 µΩ. This shunt (and the one to follow) requires a separate fixture to make proper current connections. The Special Note at the end of this document explains the fixture and physics in more detail. Figure 5  shows a Bode plot of this shunt’s impedance spectrum.

Low-Impedance-EIS-at-its-Limits-Ref30k-Booster.pdf 4.jpg

Figure 5 –Impedance spectrum of 25 µΩ shunt.

The flat portion of the curve corresponds to 24.82 ± 0.07 µΩ at 0° phase, representing less than 1% deviation from the rated value. All measurements up to 62 Hz are accurate to within 1% magnitude and within 2° phase. Performing a background subtraction of a shorted-lead spectrum reduces the magnitude error by 50% and increases the 1%/2° performance up to 100 Hz.

16.67 µΩ Shunt

The third shunt tested was rated for 50 mV at 3000 A, or 16.67 ± 0.04 µΩ. Figure 6 shows a Bode plot of this shunt’s EIS spectrum.

Low-Impedance-EIS-at-its-Limits-Ref30k-Booster

Figure 6 – Impedance spectrum of 16.67 µΩ shunt.

The flat portion of the curve corresponds to 16.60 ± 0.05 µΩ at 0° phase, representing less than 1% deviation from the rated value. All measurements up to 62 Hz are accurate to within 1% magnitude and within 2° phase. Performing a background subtraction of a shorted-lead spectrum reduces the magnitude error by a small amount but increases the 1%/2° performance up to 100 Hz.

Shunts and Background

One factor in the EIS300/Reference 3000/Reference 30k Booster system’s impressive performance is its extremely low background noise. Figure 7 shows the impedance modulus of the three shunts along with the measured background (shorted-lead spectrum) of the same system.

Low-Impedance-EIS-at-its-Limits-Ref30k-Booster.pdf 6.jpg


Figure 7 – Impedance modulus of three shunts [125, 25, and 16.67] µΩ and the shorted-lead background.

Measurement of Real Devices

As mentioned in the introduction, very low impedance is of greatest importance in the energy sector. Because ESR directly leads to power loss, minimizing ESR is critical to device performance. This becomes even more important as the voltage of the device decreases or the operating current increases. 

Figure 8 shows Nyquist plots of impedance spectra for two current technology devices (a 45 Ah Li-ion battery and a 650 F super-capacitor) and for the shorted-lead spectrum. (Note: this shorted-lead spectrum was collected using 3 Arms AC stimulation).

Low-Impedance-EIS-at-its-Limits Ref30k Booster

Figure 8 – Nyquist plots for a 650 F super-capacitor, a 45 Ah Li+ battery, and system background.

What looks like a single point at the origin is actually the complete shorted-lead spectrum over the same frequency range as those of the two devices. Not only do these devices have ESR below 1 mΩ, the impedance over most of their EIS spectra is below 1 mΩ. Next-generation devices should exhibit even lower impedance. It is critical to be able to measure these very low impedances with confidence. 

Special Note on Shunt Connections

No matter how good an impedance system is, below 1 mΩ experimental setup becomes critical. Use of twisted pairs and 4-terminal connections is mandatory.

Below 100 µΩ, however, current pathways relative to the sense leads also make a difference. Devices calibrated using a specific geometry (like shunts) make this apparent.

While the higher-resistance shunts were designed for a simple 4-probe connection that can be easily made with most potentiostats, the lower-impedance shunts needed a special setup to ensure proper current distribution. At these total impedance levels, the fractional impedance of the brass-block portion of the shunt affects how current is distributed and the measured voltage drop. 

Consider the diagram of the 16.67 µΩ shunt in Figure 9. It has four bolt holes in heavy brass blocks where high current connections are made. These holes are labeled A through D on the diagram. 

schematic shunts


Figure 9. Schematic of 16.67 and 25 µΩ shunts.

This shunt was designed to measure current flowing from the block with the AB pair to the block with the CD pair. The shunt was calibrated with connections using all four bolts. 

Ideally, all of the shunts’ resistances are in the metal with a low temperature coefficient (shown in gray on the diagram). To achieve this, the resistance of the brass blocks must be negligible. We found that this is not the case.    

We recorded three impedance spectra on the 16.67 µΩ shunt with the current carrying leads:

  • Connected to holes A and C,
  • Connected to holes B and D, and
  • Connected to a fixture that spread the current evenly through the brass blocks. 

Figure 10 shows the resistive portion of the spectra.   The data recorded with the even current distribution were close to the nominal value for the shunt. The data with the other connections had errors as large as 4%. 

Low-Impedance-EIS-at-its-Limits-Ref30k-Booster.pdf 9.jpg

Figure 10. Resistive portion of the EIS spectra, using a low-resistance shunt and different connections. Blue is A–C, red is B–D, gray is AB–CD.

The fixture we used is presented in Figure 11. We used four copper plates to spread the current through the brass blocks.  

4 terminal measurements on shunt

Figure 11. Fixture and setup for appropriate current distribution and accurate 4-terminal measurements on a shunt.

Conclusion

Using an appropriate (cell) setup and Gamry Instruments’ Reference 3000 with the Reference 30k Booster, you can make accurate EIS measurements as low as 16.67 µW.