EIS on Coatings using the FAS2 Femtostat with the ECM8 Electrochemical Multiplexer 

 


Introduction

EIS is becoming a well-accepted and widely used technique for evaluating the corrosion resistance of coatings applied to metals. It offers a number of advantages when compared to the older exposure tests:

    • EIS can detect failures before they become visible -- decreasing test time,
    • EIS gives a numerical result that is independent of the tester’s subjective judgment,
    • EIS can give information about the failure mechanism.

EIS tests are inherently non-destructive, so repeated tests can follow the degradation of a coating with exposure time. Analysis of the changes in the EIS spectrum over time often yields better information than a single measurement after a set exposure time. The time series can improve the accuracy of the test, increase the relevance of the information obtained and decrease the exposure time necessary to produce a detectable failure.

Even with the availability of high performance, lower cost EIS instrumentation, an EIS instrument generally cannot be dedicated to a single painted sample. Instead, multiple samples are exposed to a corrosive environment and periodically connected to the EIS instrument and tested. Unfortunately, each test takes some time to set up and a significant time to run, making this type of test program technician time intensive.

An Electrochemical Multiplexer, such as the Gamry Instruments’ ECM8, can help lower the labor requirements and instrumentation costs of an EIS test program. The ECM8 allows 8 samples to be connected to a single EIS instrument. An EIS spectrum can be measured on each sample in a sequential (not simultaneous) manner.

The software controlling the ECM8 will record an EIS spectrum on each sample connected to it. An extension of the software allows repetition of this set of measurements at set time intervals. In the former case, technician time is lowered because cell connections are all made at once and software setup is only done once for all eight samples. In the latter case, periodic EIS measurements are recorded over a long time period with no operator intervention at all.

The ECM8 Electrochemical Multiplexer was originally developed for use with DC corrosion measurements of bare metals. It can be used in an EIS system testing coated metal specimens with some limitations (caused by the higher sensitivity required for EIS measurements). This Note describes these limitations and offers suggestions for optimal use of the ECM8 with the Gamry Instruments EIS300 EIS Software operating with an FAS2 Femtostat.

 

EIS Background

Gamry Instruments’ Web site contains both a primer on EIS measurements and a discussion of EIS on coatings. The information in these application notes will not be repeated here.

 

Experimental Techniques

This application note includes some EIS spectra recorded on a dummy cell connected to an ECM8. These spectra were recorded using a Gamry EIS system equipped with an FAS2 Femtostat. Both the FAS2 and ECM8 were unmodified instruments randomly pulled from Gamry’s stock.

The dummy cell was enclosed in a Faraday shield, which was in turn connected to both earth ground and to the floating ground terminal on the PCI4 cell cable.

The FAS2 was connected to the ECM8 with a standard ECM8 to potentiostat cable (P/N 985-13) which was in turn connected to an FAS2 to ECM8 Cable Adapter (P/N 985-00062).  The adapter connected to the three FAS2 cell connections and also connected the FAS2 Floating Ground terminal to the ECM8's Cell Connector ground.

The ECM8 was connected to the dummy cells using standard PCI4 cell cables. Only the Working, Reference and Counter electrodes leads were connected to the cell. Unused leads were connected to the Faraday cage.

The dummy cell consisted of a dipped mica capacitor (with ± 10% tolerance) in parallel with a carbon film, high Meg resistor (± 10% tolerance). In some experiments, the spectrum of an open lead dummy cell was measured. In this case, the reference and counter electrode leads were shorted together and the working electrode lead was left dangling in the air.

In all cases, the dummy cell’s spectrum was recorded using the EIS300’s standard Potentiostatic EIS experiment. Prior to recording the spectrum, an ECM8 channel was selected by ABORTing the MUXEISP script when it had switched the ECM8 to the channel of interest. In some cases, modified scripts were used as described in the Results and Discussion section.

All spectra were recorded at a DC potential of zero volts versus the reference electrode potential.

Results and Discussion

FAS2 Baseline Data – No ECM8

An FAS2 without an ECM8 is specified to measure 15 pF in parallel with 1011 ohms with errors of less than 5% in magnitude and 10° of phase. A stock FAS2 was used to record the spectrum of a parallel RC dummy cell built using these values. The resulting Bode plot is seen in Figure 1. The FAS2 performance exceeds its specifications.
   

Figure 1
Baseline Spectrum of Dummy Cell

img_1.gif (36714 bytes)

 

The open lead Bode plot recorded using the same FAS2 is seen in Figure 2. In this figure, the line is the result of a visual fit to a parallel RC model. The equivalent capacitance of this open lead curve is less than 1 pF and the limiting resistance is greater than 8 TW. The loss of phase information at low frequency/high impedance is typical of FAS2 open lead spectra.

Figure 2
Base Line – Open Lead Curve

img_2.gif (36534 bytes)

 

FAS2 Connected to ECM8 – Stock Script and ECM8 Jumpers

The same FAS2 was then connected to an ECM8. The next test attempted to record the spectrum of the dummy cell using this setup. The resulting Bode plot is seen in Figure 3.

Comparing Figure 1 and Figure 3, you can see that the spectrum recorded though the ECM8 was very poor indeed. The magnitude should approach 1011 W at 100 mHz. In Figure 5, the magnitude at 0.1 Hz is only about 109 W.
   

Figure 3
Stock ECM Based System – Dummy Cell

img_3gif.gif (34808 bytes)

 

The Lissajous figures shown while the spectrum in Figure 3 was being recorded provided the clue used to diagnose the problem. A typical "bad" Lissajous figure is seen in Figure 4. Large spikes in the current occur as the voltage crosses zero. The charge in one of these spikes is about 10 pC.

The local potentiostat in the ECM8 is left in an "open loop" condition when it is not in use. Its output is at positive or negative saturation (about +12.5 volts or -12.5 volts). This output is not directly connected to the cell, but can couple to the cell through the capacitance of open relays and adjacent PC board traces.

As the potentiostat’s input voltage crosses the local potentiostat’s set point voltage (which is zero volts by default), the local potentiostat’s output abruptly swings from one saturation point to the other. The swing is thus about 25 volts. This voltage change will couple 10 pC of charge through a 0.4 pF capacitor. 0.4 pF is a reasonable value for stray coupling capacitance in the ECM8.
   

Figure 4
Bad Lissajous Figure Recorded with Stock System

img_4.gif (26931 bytes)

 

FAS2 Connected to an ECM8 – Jumpered to Disable the Local Pstats

The local potentiostats in the ECM8 can be disconnected from the cell by moving internal jumpers. Moving the jumpers disconnects the potentiostats’ inputs from the cells’ reference electrode inputs.

When the dummy cell experiment was repeated using an ECM8 with a disabled local potentiostat, a much better spectrum was recorded (most of the time). Figure 5 is an example. Note that the fit reports a high capacitor value and a low resistor value.

However, in some cases, distorted Lissajous figures and very noisy spectra were seen even though the local potentiostats were disabled. I believe that the disconnected potentiostat input was drifting and picking up noise, resulting in local potentiostat output swings and coupling of noise into the cell.
   

Figure 5
EIS Spectrum of Dummy Cell with Local Potentiostats Disabled

img_5.gif (31988 bytes)

 

Open Lead Curve with the Local Potentiostat Voltage at +5 Volts.

A modified MUXEISP script was generated to help eliminate the drift and pickup problems mentioned above. This script changed the set point voltage of all local potentiostats to +5 volts (in place of the default setting of zero volts). As before, ABORT was selected to halt this script on a desired ECM8 channel.

With the local potentiostats still disabled, use of this script this script eliminated the occasional glitch seen in the previous test. Spectra of similar to that in Figure 5 were recorded on the dummy cell.

Making this set point change and leaving the local potentiostats enabled also led to good quality spectra, but resulted in higher DC background currents.

Figure 6 shows an open lead spectrum recorded with the local potentiostats both biased at +5 volts and disabled.  Compare this spectrum with Figure 2. Connecting the ECM8 increases the open lead capacitance to more than 5 pF. Without the ECM8, the open lead capacitance was less than 0.5 pF. Looking back at Figure 5, the reported capacitance is about 5 pF too high - most likely the result of the open lead capacitance adding to the cell’s capacitance.
    

Figure 6
Open Lead Spectrum – No Local Pstat and 5V Potentiostat Setting

img_6.gif (32138 bytes)

 

The low frequency region of the open lead curve in Figure 6 does not fit the parallel RC model very well. The following discussion is an attempt to explain this phenomena.

The impedance at frequencies below 1 Hz was in excess of 1011 W. The AC current through 1011 W with a 10 mV excitation voltage is 100 fA rms.

The DC background current on this ECM8 channel was much larger than this AC signal. It was generally between 40 and 80 pA. This background current drifts substantially with time and temperature. The open lead curve was recorded overnight. The ambient temperature in the building fell substantially (by at least 4 ° C) while the spectrum was recorded. The DC background current also fell, because it is highly temperature dependent. Distorted Lissajous figures are observed when low frequency data was recorded on top of a shifting background. I believe that this effect caused the poor low frequency performance in the open lead experiment.

Conclusions

Script and Jumper Setting Changes

The Gamry EIS300 with the default ECM8 jumper settings and standard MUXEISP script has significant problems measuring the EIS spectra of high impedance samples.

Both disabling the ECM8’s local potentiostats and modifying the MUXEISP script to move the local potentiostat set point away from zero substantially improve the system’s performance. We recommend that both changes be made in all systems. Note that the local potentiostat is not required or used in most EIS coatings tests.

The ECM8 Operator’s Manual contains instructions telling you how to disable the ECM8’s local potentiostats. The set point of the ECM8’s local potentiostat’s is set to 5 volts by adding the following line to the MUXEISP script just prior to the loop in which data is taken:

Mux.SetDac(NIL,5.0)

This change should already be incorporated in all revisions of the EIS300 later than revision 3.03.

Typical Accuracy Limits

The ECM8 characteristics dominate the performance of the FAS2/ECM8 based EIS system.  The FAS2 performance is degraded by at least one order of magnitude when it has to contend with the added leakage currents and capacitance on the ECM8 and long interconnecting cables.

Conservatively extrapolating from the open lead spectrum, the typical accuracy contour map will be no worse than that in Figure 7.  Note that EIS through an ECM8 is not recommended for coatings with a capacitance less than 60 pF or a low frequency resistance of greater than 5 x 109 ohms.  System performance is similar to that of older, single channel systems built around Gamry's PC3 Potentiostat or EG&G's Model 273.  Tests done for this Application Note were all performed with the FAS2 Femtostat.  Similar effects were obtained with a PCI4 Potentiostat.  With either potentiostat, the ECM8 characteristics dominate the accuracy limits, since the potentiostat’s solo performance is at least two times better than that in an ECM8 system.

Note - Gamry Instruments cannot guarantee this "typical" performance.  ECM8 channels exhibit significant variation in leakage currents both within a specific ECM8 and when comparing ECM8s.   However, a second FAS2 /ECM8 measuring an 82 pF capacitor (in the 10% and 10° region of Figure 7, gave performance closer to 4% and 3° up to a DC limit of 20 G ohms.

Figure 7
Typical Accuracy Contour Map – EIS System with ECM8

img_7.gif (3427 bytes)

 

  Return to TOC

 

 

Home | Products | App Notes | Sales | Contact | News | Support | Search

Batteries | Fuel Cells | Corrosion | Paints & Coatings | Physical Electrochemistry

Gamry Instruments © 1997-2004
  
Last revised on Monday, August 21, 2006