We were recently bringing some new engineering and production employees up-to-speed on electrochemistry, since they had never seen an electrochemical cell in action. So, we quickly threw together a PTC1 Paint Test Cell with a PortHole Sample Mask (3cm2) on a mild steel (??) sample as the working electrode and filled it with some KCl solution that was handy. We think that the concentration was about 0.5M, but no one knows for sure. By the time the demonstration was over, we realized that we had measured the corrosion rate of the mild steel by three different techniques; Polarization Resistance, EIS, and our newest technique, Electrochemical Frequency Modulation, or EFM for short. If we had known that the results were
going to be so good, we would have taken more care in setting up the
experiment! Nonetheless, what follows are the results of that hastily
thrown together demo. |
||||||||||||||||||||||
|
|
Polarization Resistance |
|||||||||||||||||||||
|
|
||||||||||||||||||||||
|
|
Electrochemical Impedance Spectroscopy |
|||||||||||||||||||||
|
One way to check for the validity of a simple LPR experiment is to make an EIS measurement. The EIS measurement can also give an estimate of the solution resistance. A routine EIS measurement was performed at the Open Circuit Potential. The results of the EIS measurement are shown below. The red lines are the fit to the fairly simple model which is also shown. The inductor is needed to explain the high frequency phase shift. As in many systems, the double layer capacitance is best modeled by a constant phase element. The value of the polarization resistance, 674 +/- 13 ohm, agrees quite well with that from the LPR experiment! The uncompensated resistance (also known as the solution resistance) is quite small -- less than 1 ohm.
|
||||||||||||||||||||||
|
|
Electrochemical Frequency Modulation (EFM) |
|||||||||||||||||||||
|
Finally, we ran the same sample using the EFM technique using frequencies of 20 mHz and 50 mHz. Both of these frequencies are on the low frequency plateau on the EIS plot, above. This is required by the EFM theory.
The EFM corrosion rate (7.40 mpy) is in pretty good agreement with the rate (5.77) we calculated from the LPR measurement. However, the LPR calculation assumed that both betas were 120 mV/decade. Now that we have estimates for the Tafel constants from the EFM measurement, we can recalculate the LPR corrosion rate. The table, below, summarizes the data from all three experiments. The agreement between the three non-destructive techniques is excellent! Not bad for a lunch-time
exercise!
You can read about another, more careful, comparison done at the 2004 Corrosion Short Course at Penn State. |
||||||||||||||||||||||
|